• Category list

Products£­Industrial temperature sensor£­Assembling type RTDs£­Inspection standards

Inspection Standard of RTDs

Appearance and Construction

Inspect

Terminal head, welded and/or soldered areas, screwed areas, name plate and other parts shall be inspected with naked eyes. Dye penetrating check or airtight test shall be performed on welded and/or soldered areas, if required.

 

Dimensional

Inspection

OD of protection tube

Φ3.0¡«Φ5.0 mm ±0.05 mm

Φ6.0¡«Φ6.4 mm ±0.06 mm

Φ8¡«Φ22mm ±1%

 

Length of protection tube

<150 mm ± 2.0 mm

≥150 mm ± 1.5%

 

Nipple length

±3 mm

 

Other parts

Dimensions of the parts specified on the drawing shall be measured with a measuring tool. Unless otherwise specified, the measurement shall be performed in accordance with IEC.

 

Nominal

resistance test

Nominal resistance value

0¡æ       Icing point of water

 

Insulation

Resistance test

The resistance between the conductor and the sheath is measured at room temperature.

 

OD of protection tube

Characteristics

 

Φ3.0¡«Φ22

100 MΩ / 100 VDC

 

Qualified Certificate

Inspection Certificate is attached to the product. Whenever the inspection results are especially required, please so advise.

¡¡
ÔÚÏ߿ͷþ